O2 |
0 … 21 % |
NO(x) |
0 … 1.000 ppm* overload up to 5.000 ppm (option) |
CO (H2 compensated) |
0 … 10.000 ppm* overload up to 20.000 ppm (option) |
Differential pressure |
at basic unit ± 100hPa, at remote control ± 200hPa |
External pressure sensor |
up to 40 bar (option) |
Temperature |
up to 800 °C or up to 1.100 °C (with suitable sensor) |
Fluegas temperature |
0 … 800°C (stainless steel tube) 0 … 1.100°C (inconeltube) |
Relative humidity |
3 … 98 % |
Amount of leakage |
0 … 8 l/h |
External temperature sensor |
K-type standard |
|
|
Calculated values |
|
Dew point |
calculated °C |
Losses |
0 …99,9 % |
|
|
Dimensions and other data |
Gas conditioning |
illuminated condensate trap, alternatively gas cooler |
Sample gas flow monitoring |
optional, standard with soot measurement |
Flow velocity |
with pitot tube 1 … 40 with vane 0,25 … 35 m/s |
Soot measurement |
optionally available as built-in feature |
Operating temperature |
+ 5 … + 45 °C, max. 95 % RF, non condensing |
Power supply |
grid 100 … 240 Vac / 50 … 60 Hz, 1,4 A, 12 V DC/5A |
Battery capacity |
basic unit up to 10 hours resp. with gas cooler up to 6 hours / remote control up to 30 hours |
Weight |
7,4 kg with 2 sensors, probe, power supply, case |
Dimensions |
470x314x235 mm (HxWxD) |
TÜV approval |
By RgG 296 |
4/8 Pa-Test |
optionally available |
Data storage |
internally 1MB or externally 2 GB |
Data transfer |
via USB cable, optionally via Bluetooth |
PC-interface |
mini-USB, optionally Bluetooth |
Display |
3,5″ TFT colour display |
Pressure ranges |
internally up to 1,5 bar (option) externally up to 40 bar (option) |